61
ALARM PARAMETERS
Setting temperature alarm
When the temperature goes out of the preset range (the set chamber temperature ± the set temperature
alarm value), the alarm activates. The temperature alarm value can be set as follows.
Settable range: ±1.0
o
C – ±5.0
o
C
Factory setting: ±1.0
o
C
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display changes to the Menu screen.
2. On the Menu screen, move the cursor to Alarms & Controls [Fig. 1]
using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display changes to the Alarms & Controls screen.
3. When the cursor in the Alarms & Controls screen is on Temp/Gas
Alarm Set [Fig. 2], press the enter key (ENTER).
▶
The display changes to the Temp/Gas Alarm Set screen.
4. When the cursor in the Temp/Gas Alarm Set screen is on Temp Alarm
[Fig. 3], press the enter key (ENTER).
▶
The right side of the display changes to the Temp Alarm screen, and
the current setting value (±1.0) is displayed [Fig. 4].
5. Move the cursor using the left/right keys (
◁▷
) and select the digit you
want to change, and then use the up/down keys (
△▽
) to change the
number.
6. Press the enter key (ENTER).
▶
The value is confirmed and the display returns to the Temp/Gas Alarm
Set screen.
*If the entered setting value is out of the settable range, the following
message is displayed on the right side of the display. Press any key on
the control panel to change to the Temp Alarm screen [Fig. 4].
7. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
<Left side of the display> [Fig. 2]
Alarms & Controls
Temp/Gas Alarm Set
Other Alarm Set
UV Setting
Gas Supply Setting
Date & Time
Keypad Click Set
▼
Temp Alarm
Invalid Value
Press Any Key
±1.0°C – ±5.0°C
<Right side of the display>
<Right side of the display> [Fig. 4]
Temp Alarm
±1.0
°C
±1.0°C – ±5.0°C
<Left side of the display> [Fig. 1]
Menu
Set Value
Data Log
Lock
Alarms & Controls
<Left side of the display> [Fig. 3]
Temp/Gas Alarm Set
Temp Alarm
Temp Alarm Delay
CO
2
Alarm
CO
2
Alarm Delay
O
2
Alarm
O
2
Alarm Delay
Summary of Contents for MCO-50ML
Page 103: ...103 MEMO ...