Troubleshooting
131
Action: Check for an obstruction in the sample line.
Make sure the sample line, the sample
probe and the inlet filter are not blocked.
WARNING
Do not replace the inlet filter in a hazardous location.
Action: Replace the inlet filter. See
Replacing the
Sample Inlet Filter
on page 120.
Action: Ensure the sample outlet, located on the
underside of 2020, is not obstructed.
Cause: UV lamp is too wide, causing flow to be restricted.
WARNING
Do not remove or replace the UV lamp in a hazardous location.
Action: If you have a UV lamp with a white serial
number label, it is possible that the lamp is
too wide for the lampholder. Contact the
Service Department.
Cause: 2020 has been exposed to a solvent that can pass
through the inlet filter and liquid has been aspirated.
Action: Contact the Service Department.
Cause: The pump has failed.
Action: Contact the Service Department.
Summary of Contents for Photovac 2020
Page 2: ......
Page 3: ...Photovac Model 2020 Portable Photoionization Detector...
Page 5: ...Reference No 600762 Rev E...
Page 6: ......
Page 13: ...Notices and Warnings...
Page 14: ......
Page 23: ...Introduction 1...
Page 24: ......
Page 38: ......
Page 39: ...Tutorial Session 2...
Page 40: ......
Page 51: ...Tutorial Session 49 Mode on page 62 for details of switching between display modes...
Page 60: ...2020 Software Manual 58 the datalogger or at the end of the datalogger will free usable space...
Page 67: ...Tutorial Session 65...
Page 68: ......
Page 69: ...Detailed Operation 3...
Page 70: ......
Page 88: ...2020 Software Manual 86...
Page 89: ...Connecting Accessories 4...
Page 90: ......
Page 115: ...Routine Maintenance 5...
Page 116: ......
Page 126: ...2020 Software Manual 124...
Page 127: ...Troubleshooting 6...
Page 128: ......
Page 153: ...Troubleshooting 151...
Page 154: ......
Page 155: ...Technical Description 7...
Page 156: ......
Page 163: ...Appendices 8...
Page 164: ......
Page 181: ...Appendices 179...