2020 Software Manual
72
9.
Select “Set”, “Cal” and “Span”. 2020 asks for the span
gas concentration. Enter the known span gas
concentration, without pressing the ENTER key to
confirm it.
10.
Insert the 2020 sample probe into the adapter tubing
from the regulator. See
Figure 11
.
11.
Ensure the calibration gas cylinder is upright and open
the regulator by turning the valve counter clockwise.
Open the regulator until the ball is 1/8” from its rest
position.
NOTE:
Do not set the flow rate too high.
12.
Press the ENTER key. 2020 sets its sensitivity.
13. When the display reverts to the default display, 2020 is
calibrated and ready for use.
14. Remove the adapter tubing from the inlet and close the
regulator.
If you turn off 2020 in the middle of zeroing or spanning, the next
time you turn it on it will display a Cal status. This indicates that you
need to calibrate 2020.
NOTE:
While the Cal status is active, all alarms are deactivated.
Summary of Contents for Photovac 2020
Page 2: ......
Page 3: ...Photovac Model 2020 Portable Photoionization Detector...
Page 5: ...Reference No 600762 Rev E...
Page 6: ......
Page 13: ...Notices and Warnings...
Page 14: ......
Page 23: ...Introduction 1...
Page 24: ......
Page 38: ......
Page 39: ...Tutorial Session 2...
Page 40: ......
Page 51: ...Tutorial Session 49 Mode on page 62 for details of switching between display modes...
Page 60: ...2020 Software Manual 58 the datalogger or at the end of the datalogger will free usable space...
Page 67: ...Tutorial Session 65...
Page 68: ......
Page 69: ...Detailed Operation 3...
Page 70: ......
Page 88: ...2020 Software Manual 86...
Page 89: ...Connecting Accessories 4...
Page 90: ......
Page 115: ...Routine Maintenance 5...
Page 116: ......
Page 126: ...2020 Software Manual 124...
Page 127: ...Troubleshooting 6...
Page 128: ......
Page 153: ...Troubleshooting 151...
Page 154: ......
Page 155: ...Technical Description 7...
Page 156: ......
Page 163: ...Appendices 8...
Page 164: ......
Page 181: ...Appendices 179...