2020 Software Manual
34
Charging a fully discharged battery pack will take approximately 4
hours.
Optionally you can use the off line charger (Part No. MX350015) to
charge the battery pack independently of the instrument. See
Off
Line Charger
on page 110.
If you are charging the battery pack in the instrument you can use
all the features of 2020 while the battery pack is being charged.
Leaving the AC adapter connected to 2020 will not harm the battery
or the AC adapter in any way. If 2020 is to be left indefinitely, leave
it connected to the AC adapter so that the battery will be fully
charged and ready for operation.
On average a fully charged battery pack will provide 8 hours of
continuous operation. Battery life is shorter if the instrument is
turned off and then on again repeatedly, or if the backlighting is
turned on.
Overview
2020 measures the concentration of airborne photoionizable gases
and vapors and automatically displays and records these
concentrations. It does not distinguish between individual pollutants.
The reading displayed represents the total concentration of all
photoionizable chemicals present in the sample. 2020 is factory set
to display concentration in units of ppm or mg/m³.
2020 operates automatically. The meter display updates itself once
per second. You can read concentrations directly from the meter
display. If you are using the dilution probe (Part No. MX350013)
you must multiply the displayed reading by the dilution factor. See
Dilution Probe
on page 101 for details.
2020 is always performing short term exposure limit (STEL), time
weighted average (TWA) and PEAK calculations. You can view
any of these results, but only one mode may be viewed at a time.
Summary of Contents for Photovac 2020
Page 2: ......
Page 3: ...Photovac Model 2020 Portable Photoionization Detector...
Page 5: ...Reference No 600762 Rev E...
Page 6: ......
Page 13: ...Notices and Warnings...
Page 14: ......
Page 23: ...Introduction 1...
Page 24: ......
Page 38: ......
Page 39: ...Tutorial Session 2...
Page 40: ......
Page 51: ...Tutorial Session 49 Mode on page 62 for details of switching between display modes...
Page 60: ...2020 Software Manual 58 the datalogger or at the end of the datalogger will free usable space...
Page 67: ...Tutorial Session 65...
Page 68: ......
Page 69: ...Detailed Operation 3...
Page 70: ......
Page 88: ...2020 Software Manual 86...
Page 89: ...Connecting Accessories 4...
Page 90: ......
Page 115: ...Routine Maintenance 5...
Page 116: ......
Page 126: ...2020 Software Manual 124...
Page 127: ...Troubleshooting 6...
Page 128: ......
Page 153: ...Troubleshooting 151...
Page 154: ......
Page 155: ...Technical Description 7...
Page 156: ......
Page 163: ...Appendices 8...
Page 164: ......
Page 181: ...Appendices 179...