Troubleshooting
127
General Information
If you have a service related question about 2020, consult this
manual first. If you cannot find the answer in this documentation,
contact the Service Department.
When you call, you should have your 2020 in front of you. You
should also have this manual at hand.
Lastly, please have the following information ready:
1. A description of what happened and what you were
doing when the problem occurred.
2. Any corrective action that you have tried.
3. The exact wording of any messages that appeared on
the display.
WARNING
Do not service 2020 in a hazardous location.
Fault Messages
When the “Falt” status is displayed, 2020’s operation is
compromised.
Falt1: Signal from zero gas is too high.
Cause: If another fault occurred while 2020 was setting its
zero point, then this fault is displayed.
Action: Ensure no faults are occurring and calibrate
2020 again.
Summary of Contents for Photovac 2020
Page 2: ......
Page 3: ...Photovac Model 2020 Portable Photoionization Detector...
Page 5: ...Reference No 600762 Rev E...
Page 6: ......
Page 13: ...Notices and Warnings...
Page 14: ......
Page 23: ...Introduction 1...
Page 24: ......
Page 38: ......
Page 39: ...Tutorial Session 2...
Page 40: ......
Page 51: ...Tutorial Session 49 Mode on page 62 for details of switching between display modes...
Page 60: ...2020 Software Manual 58 the datalogger or at the end of the datalogger will free usable space...
Page 67: ...Tutorial Session 65...
Page 68: ......
Page 69: ...Detailed Operation 3...
Page 70: ......
Page 88: ...2020 Software Manual 86...
Page 89: ...Connecting Accessories 4...
Page 90: ......
Page 115: ...Routine Maintenance 5...
Page 116: ......
Page 126: ...2020 Software Manual 124...
Page 127: ...Troubleshooting 6...
Page 128: ......
Page 153: ...Troubleshooting 151...
Page 154: ......
Page 155: ...Technical Description 7...
Page 156: ......
Page 163: ...Appendices 8...
Page 164: ......
Page 181: ...Appendices 179...