2020 Software Manual
48
Mode
on page 62 for details of switching between display modes.
Time Weighted Average (TWA) Mode
The TWA accumulator sums concentrations every second until 8
hours of data have been combined. If this value exceeds the TWA
alarm setting, a TWA alarm is generated. The TWA is not
calculated using a moving average. Once 8 hours of data have been
summed, the accumulation stops. In order to reset the TWA
accumulator, press the “Clr” key.
This sum will only be complete after 8 hours, so the meter displays
the current sum divided by 8 hours. While you are in TWA mode,
the time on the status display will show the number of minutes and
hours of data that TWA has accumulated. When this reaches 8:00
or 8 hours, 2020 stops accumulating data and the TWA is complete.
TWA calculations are always being performed by 2020. You can
display the results of the calculations by selecting TWA as the
Display mode. See
Summary of Contents for Photovac 2020
Page 2: ......
Page 3: ...Photovac Model 2020 Portable Photoionization Detector...
Page 5: ...Reference No 600762 Rev E...
Page 6: ......
Page 13: ...Notices and Warnings...
Page 14: ......
Page 23: ...Introduction 1...
Page 24: ......
Page 38: ......
Page 39: ...Tutorial Session 2...
Page 40: ......
Page 51: ...Tutorial Session 49 Mode on page 62 for details of switching between display modes...
Page 60: ...2020 Software Manual 58 the datalogger or at the end of the datalogger will free usable space...
Page 67: ...Tutorial Session 65...
Page 68: ......
Page 69: ...Detailed Operation 3...
Page 70: ......
Page 88: ...2020 Software Manual 86...
Page 89: ...Connecting Accessories 4...
Page 90: ......
Page 115: ...Routine Maintenance 5...
Page 116: ......
Page 126: ...2020 Software Manual 124...
Page 127: ...Troubleshooting 6...
Page 128: ......
Page 153: ...Troubleshooting 151...
Page 154: ......
Page 155: ...Technical Description 7...
Page 156: ......
Page 163: ...Appendices 8...
Page 164: ......
Page 181: ...Appendices 179...