2020 Software Manual
132
Troubleshooting
Problem: Very low or no instrument response detected, yet
compounds are known to be present.
Cause: 2020 has not been calibrated properly.
Action: Ensure the calibration gas is of a reliable
concentration and then calibrate the
instrument as outlined in Chapter 3.
After the instrument has been calibrated,
sample the bag of calibration gas. A reading
equivalent to the calibration gas should be
displayed. If not, contact the Service
Department.
WARNING
Do not remove or recharge the battery pack in a hazardous location.
Action: Disconnect the battery charger before
calibrating 2020. See Chapter 3 for
calibration instructions.
Cause: Cal Memories have not been programmed
correctly.
Action: Program all the Cal Memories you require
for your application. You must use the
correct calibration gas and concentration
for each Cal Memory. See
Programming
the Cal Memories
o n page 76.
Cause: Response factor has been set to zero.
Summary of Contents for Photovac 2020
Page 2: ......
Page 3: ...Photovac Model 2020 Portable Photoionization Detector...
Page 5: ...Reference No 600762 Rev E...
Page 6: ......
Page 13: ...Notices and Warnings...
Page 14: ......
Page 23: ...Introduction 1...
Page 24: ......
Page 38: ......
Page 39: ...Tutorial Session 2...
Page 40: ......
Page 51: ...Tutorial Session 49 Mode on page 62 for details of switching between display modes...
Page 60: ...2020 Software Manual 58 the datalogger or at the end of the datalogger will free usable space...
Page 67: ...Tutorial Session 65...
Page 68: ......
Page 69: ...Detailed Operation 3...
Page 70: ......
Page 88: ...2020 Software Manual 86...
Page 89: ...Connecting Accessories 4...
Page 90: ......
Page 115: ...Routine Maintenance 5...
Page 116: ......
Page 126: ...2020 Software Manual 124...
Page 127: ...Troubleshooting 6...
Page 128: ......
Page 153: ...Troubleshooting 151...
Page 154: ......
Page 155: ...Technical Description 7...
Page 156: ......
Page 163: ...Appendices 8...
Page 164: ......
Page 181: ...Appendices 179...