2020 Software Manual
80
Memory that must be used when monitoring the site. Also include
any reference information that will help you define the site and the
monitoring application. If you create your schedule using
spreadsheet software, you can later download 2020 data to a
computer and then copy it into the spreadsheet for further
calculations.
Once you have programmed 2020, and prepared a list of sites to be
monitored, you will move around to each location and manually log
data at each site.
To log data according to a monitoring schedule:
1.
You must determine the number of calibration standards
that will be required to perform manual monitoring for
your application. Program and calibrate all the
calibration memories that you need. See
Calibration
Using the Flow-Match Regulator
on page 70 and
Programming the Cal Memories
on page 76.
NOTE:
It does not matter which Cal Memory is selected or which
response factor is entered, 2020’s response is not specific to
any one compound. The reading displayed always represents
the total concentration of all ionizable compounds in the
sample.
2.
Ensure the 2020 is in PEAK mode. To change the
mode, press ENTER and select “Disp”. Press “Mode”
and select PEAK.
3.
Switch to manual operation. Press ENTER and then
“Log”. Select “Mode”. Use the “Next” and “Prev”
keys to scroll through the list. When Manual is displayed
press ENTER. When you switch between interval and
manual operation, the datalogger will be cleared. Press
“YES” to confirm your selection and clear the
datalogger. If you do not want to lose your previously
recorded data, press “NO”, then print or save the data
to disk before changing to manual operation. See
Summary of Contents for Photovac 2020
Page 2: ......
Page 3: ...Photovac Model 2020 Portable Photoionization Detector...
Page 5: ...Reference No 600762 Rev E...
Page 6: ......
Page 13: ...Notices and Warnings...
Page 14: ......
Page 23: ...Introduction 1...
Page 24: ......
Page 38: ......
Page 39: ...Tutorial Session 2...
Page 40: ......
Page 51: ...Tutorial Session 49 Mode on page 62 for details of switching between display modes...
Page 60: ...2020 Software Manual 58 the datalogger or at the end of the datalogger will free usable space...
Page 67: ...Tutorial Session 65...
Page 68: ......
Page 69: ...Detailed Operation 3...
Page 70: ......
Page 88: ...2020 Software Manual 86...
Page 89: ...Connecting Accessories 4...
Page 90: ......
Page 115: ...Routine Maintenance 5...
Page 116: ......
Page 126: ...2020 Software Manual 124...
Page 127: ...Troubleshooting 6...
Page 128: ......
Page 153: ...Troubleshooting 151...
Page 154: ......
Page 155: ...Technical Description 7...
Page 156: ......
Page 163: ...Appendices 8...
Page 164: ......
Page 181: ...Appendices 179...