7.
Use the larger of the two hexagonal wrenches supplied (3/16 inch) to unlock
the main cover of the instrument by turning the locking screw
counterclockwise until it is loose.
The lid of the Spectrum RX 1 creeps up approximately 2 cm
(about 15 turns of the wrench).
WARNING
Make sure that the power cable is not connected.
8.
Lift the main cover from the front.
9.
Support the cover with the prop, which is stored on the underside of the
main cover to the right. Insert the free end into the notched receptacle
(Figure 2).
Do not support the prop on the power inlet module, which is behind the
notched receptacle.
Prop
Prop storage clip
Notched
receptacle
Power inlet module
Figure 2 Propping Open the Main Cover