Spectrum RX I Getting Started Guide . 41
2.
Remove the two round inserts from the left-hand side of the top of the
main cover (Figure 17). To do this, push down the edge of the insert that
is near the back of the Spectrum RX I.
The front of the insert lifts up.
Figure 17 Removing the Inserts
3.
Lift away the insert.
A hole is visible below each insert.
4.
Insert the smaller hexagonal wrench down the center of the left-hand
hole (the handle of the wrench will be at a slight angle towards the left
front corner of the instrument).
5.
Push the wrench down onto the top of the adjustment screw. Gently
rotate the wrench until it drops into the top of the adjustment screw.
NOTE:
Scan errors may be caused when the wrench is first engaged. If the
scan status box changes to red with the word
ERROR,
wait until the
scan has recovered before proceeding. This may take several seconds,
and the scan status box will return to
MONITOR
when the instrument
is ready to continue.
6.
Carefully adjust the screw until the energy bar on the screen is within one
whole number of the maximum obtainable.