Spectrum RX I Getting Started Guide . 29
Hexagonal
wrench
Septum
cover
Figure 9 Use of the Hexagonal Wrench
1.
Insert the smaller hexagonal wrench through the left-hand septum cover
such that the handle of the wrench is at an angle of about 6° towards the
left front corner of the instrument.
2.
Push the wrench down onto the top of the adjustment screw (a depth of
about 4 cm) as shown in Figure 9. Gently rotate the wrench until it drops
into the top of the adjustment screw.
3.
Turn the screw a quarter-turn clockwise and then a quarter-turn
counterclockwise to familiarize yourself with the feel of turning the screw.
4.
Remove the hexagonal wrench from the left-hand septum cover and
insert it into the right-hand septum cover such that the handle of the
wrench is at an angle of about 6° towards the left front corner of the
instrument.