47
Installation
Hot acid vapors may be released.
The supplied vent tubes must therefore always be properly connected and
attached to an exhaust system.
Make sure the PFA tube is securely attached so that it cannot pop off its
connection fitting in case of a percussive pressure burst as occurs when a
rupture disc activates.
Note
The laboratory is responsible for ensuring that the exhaust system can
appropriately vent the released gases. Make sure that all ventilation
procedures conform to applicable laws.
Note
Important Note for North American Users:
Voltage of mains supply has to be in the range of 230V ± 15V AC. Even
with a maximum load of approximately 2700 W the supply-voltage should
not drop below 215 V AC. Low voltage can result in low microwave power
delivery, excessive magnetron or inverter heating and erratic temperature
and pressure measurements.
!
Warning
Summary of Contents for TITAN MPS
Page 1: ...TITAN MPS Customer Users Guide ATOMIC SPECTROSCOPY ...
Page 2: ......
Page 3: ...Titan MPS Users Guide ...
Page 55: ...45 Power Cord Options Buck and Boost Transformer 0 5 KVA Japan and U S Canada Connections ...
Page 98: ...88 Operation Figure 5 27 Sample Details review 1 8 Figure 5 28 Sample Details review 9 16 ...
Page 104: ...94 Operation Sample Details Entering individual sample information ...
Page 105: ...95 Digestion Results Digestion Results Reviewing the data from completed digestions ...
Page 109: ...99 Data Management Data Management Options available via the Data button on the Home screen ...
Page 120: ...110 Maintenance ...
Page 122: ...112 Maintenance ...
Page 132: ...122 Appendix Login Login 1 Select the login access desired ...
Page 133: ...123 Index of Controller Screens Login Passcode 1 Enter the login passcode ...
Page 156: ...146 Appendix ...
Page 160: ......