Preparing Your
Laboratory
2
Introduction
The Titan MPS has some special requirements that you must consider when setting
up the laboratory. The laboratory must fulfill these requirements, described in this
chapter, before you can operate the microwave correctly.
Environmental Requirements
The location where the Titan MPS is to be installed must fulfill the criteria listed
below:
•
Indoors.
•
Free from vibrations.
•
Free from dust, smoke and corrosive fumes.
•
Out of direct sunlight.
•
Away from radiators and heaters.
•
Have a suitable fume ventilation system.
•
Do not use above 2000 m in elevation.
•
Ambient room temperature b15 °C to + 35 °C (59-95 °F), In addition,
the temperature should lie between -40°C (-40°F) and +70°C (158°F) during
shipping and storage.
•
Relative humidity between 20 and 85%, non-condensing.
Summary of Contents for TITAN MPS
Page 1: ...TITAN MPS Customer Users Guide ATOMIC SPECTROSCOPY ...
Page 2: ......
Page 3: ...Titan MPS Users Guide ...
Page 55: ...45 Power Cord Options Buck and Boost Transformer 0 5 KVA Japan and U S Canada Connections ...
Page 98: ...88 Operation Figure 5 27 Sample Details review 1 8 Figure 5 28 Sample Details review 9 16 ...
Page 104: ...94 Operation Sample Details Entering individual sample information ...
Page 105: ...95 Digestion Results Digestion Results Reviewing the data from completed digestions ...
Page 109: ...99 Data Management Data Management Options available via the Data button on the Home screen ...
Page 120: ...110 Maintenance ...
Page 122: ...112 Maintenance ...
Page 132: ...122 Appendix Login Login 1 Select the login access desired ...
Page 133: ...123 Index of Controller Screens Login Passcode 1 Enter the login passcode ...
Page 156: ...146 Appendix ...
Page 160: ......