102
Maintenance
Oven Chamber
The oven should be wiped out with a moist cloth at regular intervals as well as
whenever a rupture disc vents. All relevant safety measures should be observed for
the acids used. Careful attention should be used when dealing with hydrofluoric
acid.
Allow the oven to thoroughly dry before again using it!
Display
Use only a dry cloth to wipe off the display.
PFA Gas Containment Manifold Exhaust Tube
After a rupture disc vents the PFA exhaust tube will be contaminated with acid. To
disassemble the tube unscrew the fittings on both ends. Flush the tube with water
and dry with pressurized air. Reattach the tube using the screw fittings, ensuring a
tight seal.
Vessels
Since the vessels exhibit practically no memory effects it is generally sufficient to
thoroughly rinse out the vessels with distilled water after each digestion procedure
that results in complete sample digestion.
If discoloration is noted or if carry over is suspected, the vessels are cleaned by a
digestion using clean acid with no sample. The Application Notebook has the
vessel cleaning method included.
When running samples that have large differences in analyte concentrations it is
advisable to perform a cleaning digestion prior to of the next set of digestions. (for
example when the analyte concentration changes from % to ppm, ppb, or b ppb).
Cleaning Methods
The necessary cleaning method depends on the degree of the previously performed
digestion and its further use.
Summary of Contents for TITAN MPS
Page 1: ...TITAN MPS Customer Users Guide ATOMIC SPECTROSCOPY ...
Page 2: ......
Page 3: ...Titan MPS Users Guide ...
Page 55: ...45 Power Cord Options Buck and Boost Transformer 0 5 KVA Japan and U S Canada Connections ...
Page 98: ...88 Operation Figure 5 27 Sample Details review 1 8 Figure 5 28 Sample Details review 9 16 ...
Page 104: ...94 Operation Sample Details Entering individual sample information ...
Page 105: ...95 Digestion Results Digestion Results Reviewing the data from completed digestions ...
Page 109: ...99 Data Management Data Management Options available via the Data button on the Home screen ...
Page 120: ...110 Maintenance ...
Page 122: ...112 Maintenance ...
Page 132: ...122 Appendix Login Login 1 Select the login access desired ...
Page 133: ...123 Index of Controller Screens Login Passcode 1 Enter the login passcode ...
Page 156: ...146 Appendix ...
Page 160: ......