![Perkin-Elmer TITAN MPS User Manual Download Page 125](http://html1.mh-extra.com/html/perkin-elmer/titan-mps/titan-mps_user-manual_1530159125.webp)
115
Microwave Oven
Electrical breaker on the
power switch triggers fre-
quently.
Magnetron defective.
Short circuit in the device.
Contact your PerkinElmer
Service Representative.
RCCB switch of the supply
line triggers often.
Magnetron defective.
Short circuit in the device.
Contact your PerkinElmer
Service Representative.
Lack of sample pressure or
samples not reaching temper-
ature.
Vessel seals not sealing effec-
tively.
Use the seal lip-forming tool
each time to ensure a cor-
rectly formed seal. Replace
old or damaged seals.
Problem
Possible Cause(s)
Corrective Action
Summary of Contents for TITAN MPS
Page 1: ...TITAN MPS Customer Users Guide ATOMIC SPECTROSCOPY ...
Page 2: ......
Page 3: ...Titan MPS Users Guide ...
Page 55: ...45 Power Cord Options Buck and Boost Transformer 0 5 KVA Japan and U S Canada Connections ...
Page 98: ...88 Operation Figure 5 27 Sample Details review 1 8 Figure 5 28 Sample Details review 9 16 ...
Page 104: ...94 Operation Sample Details Entering individual sample information ...
Page 105: ...95 Digestion Results Digestion Results Reviewing the data from completed digestions ...
Page 109: ...99 Data Management Data Management Options available via the Data button on the Home screen ...
Page 120: ...110 Maintenance ...
Page 122: ...112 Maintenance ...
Page 132: ...122 Appendix Login Login 1 Select the login access desired ...
Page 133: ...123 Index of Controller Screens Login Passcode 1 Enter the login passcode ...
Page 156: ...146 Appendix ...
Page 160: ......