9
Precautions
All pressure vessels are equipped with a sealed connection to a gas collection
system which, in turn, must be connected to an exhaust system via PFA tubing. The
PFA tubing is connected to the screw-on connection located in the center of the top
of the microwave. Any hot acid vapors released when the rupture disc activates are
safely removed through this PFA vent tubing.
Temperature Control
The system is equipped with Direct Temperature Control (DTC) of all sample
temperatures as described in the section,
Temperature Monitoring on page 33
. An
emergency shut-off occurs in case of a rupture disc break.
Pressure Control
The system is also equipped with a pressure monitoring system for a reference
vessel located at position one as described in the
Repairs/Customer Service on
page 107
.
Over Temperature Protection
For safety reasons, the microwave automatically switches itself off in case the
magnetron, the microwave oven, or the high voltage transformer overheats.
Precautions
Be sure to read and understand the precautions listed below. It is also
recommended to post a copy of the precautions near or on the instrument
shelf.
Caution
The protection provided by this instrument may be impaired if the
instrument is used in a manner not specified by PerkinElmer.
!
Warning
Summary of Contents for TITAN MPS
Page 1: ...TITAN MPS Customer Users Guide ATOMIC SPECTROSCOPY ...
Page 2: ......
Page 3: ...Titan MPS Users Guide ...
Page 55: ...45 Power Cord Options Buck and Boost Transformer 0 5 KVA Japan and U S Canada Connections ...
Page 98: ...88 Operation Figure 5 27 Sample Details review 1 8 Figure 5 28 Sample Details review 9 16 ...
Page 104: ...94 Operation Sample Details Entering individual sample information ...
Page 105: ...95 Digestion Results Digestion Results Reviewing the data from completed digestions ...
Page 109: ...99 Data Management Data Management Options available via the Data button on the Home screen ...
Page 120: ...110 Maintenance ...
Page 122: ...112 Maintenance ...
Page 132: ...122 Appendix Login Login 1 Select the login access desired ...
Page 133: ...123 Index of Controller Screens Login Passcode 1 Enter the login passcode ...
Page 156: ...146 Appendix ...
Page 160: ......