30
The table shows a measurement per line and consists of the following columns:
•
measurement identification number;
•
“A”: identification of the logical electrode that will play role A in the
measurement, cannot be changed by the operator;
•
“B”: identification of the logical electrode that will play role B in the
measurement, cannot be changed by the operator;
•
“M”: identification of the logical electrode that will play role M in the
measurement, cannot be changed by the operator;
•
“N”: identification of the logical electrode that will play role N in the
measurement, cannot be changed by the operator;
•
“K”: geometric coefficient value calculated using the role information A, B, M
and N of the Measurement Table and the geometric position of the logic electrodes
contained in the Electrode Table, it cannot be modified by the operator.
A value of “0” in the identification field of an electrode means that no electrode is
associated with that role in that measurement.
F
IGURE
16:
M
EASUREMENTS TABLE FOR A TEMPLATE