LMX SERIES® OPERATION MANUAL
SECTION 4: Technical Specifications
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
LMX Series Power Source Operation Manual
Page 39 of 463
4.3
Overload Operation
The LMX-Series Power Source will deliver up to 150% of rated rms output current into unity
power factor loads at full-rated output voltage. Low output power factor, reduced output
voltage, and elevated ambient temperatures will increase the internal dissipation of the
power source and can cause overload shutdown due to over-temperature conditions.
Length of time to reach over-temperature varies with the models and the above parameters
4.4
AC Output – All Models
OUTPUT
SPECIFICATION
Power
Output
See Model Tables Section 4.1
Power Factor
See Model Tables Section 4.2
Voltage
Mode
AC
Direct Coupled Range
0-135 Vac LN / 0-234 Vac LL
T-Option Ranges
Turns ratios: 1:1.5, 1:2.0, 1:2.5
Programming Resolution
0.01 V
Accuracy
±0.25% F.S. / ±0.1% in CSC mode
Waveforms
(200 Max.)
Sine, Square, Triangle,
Clipped (THD), Arbitrary
DC Offset
< 20 mV
Harmonic Distortion (Vthd)
(full, resistive load, up to 360Vrms L-N)
15~1000 Hz: < 0.1%
1000~5000 Hz: < 0.25%
Output Noise
<
50 mVrms
Load Regulation
Direct Coupled: ± 0.0% (CSC Mode)
Transformer Coupled: ± 0.0% (CSC Mode)
TR 1.5:1 2% CSC Mode off
TR 2.0:1 4% CSC Mode off
TR 2.5:1 5% CSC Mode off
Line Regulation
< 0.1% for 10% Line Change
Voltage Sense
External Sense, max. voltage drop 5% F.S.
Voltage Response Time
5 μsec
typical to a step load change.
Small Signal Bandwidth
5 Hz to 40 kHz, ±3dB @ 10% of Voltage Range.
Isolation
Output Neutral to Chassis
150 Vrms Max (Direct Coupled Mode)
In Transformer Coupled mode, Neutral must be connected
to Chassis.
Frequency
Direct Coupled Range
15.00 – 5000.0 Hz
Transformer Coupled Range
45.00 – 5000.0 Hz
Programming Resolution
0.01 Hz
Accuracy
± 0.005% / 50 ppm
Current
Range
See Model Tables Section 4.2
Programming Resolution
0.01 Arms