LMX SERIES™ OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
LMX Series Power Source Operation Manual
Page 243 of 463
Query Format
PROGram:EXECuted:MEMory?
Description
This command selects the specific program memory type for storing
and retrieving programs. Default is INTERNAL memory. Alternatives
are RAM and USB devices. The memory type can also be specified as
an optional parameter with several PROGram commands where
indicated in this manual.
Parameters
Memory types: INTERNAL | RAM | USB
Parameter Format
<cr>
Example
PROG:EXEC:MEM USB
Query Format
PROGram:EXECuted:MEMory?
Returned Data Format
<cr>
Query Example
PROG:EXEC:MEM?
USB
Command Syntax
PROGram:POWOn <nr1>
Description
Returns the number of the stored program that will be recalled at
power on.
Parameters
Program number
Parameter Format
<nr1>
Example
PROG:POWO 1
Query Format
PROGram:POWOn?
Returned Data Format
<nr1>
Query Example
PROG:POWO?
1
Command Syntax
PROGram:DELete:ALL <cr>
Description
This command deletes all programs, waveforms and setup values,
performs device RESET, loads Program #1 with *RST default values,
executes MANUAL MODE with *RST values. Waveforms (1-16) are
re-loaded from internal Flash memory. Attempting to DELete an
EXECuting Program will result in an error message.
Parameters
Optional: Memory type: INTERNAL | RAM | USB
Parameter Format
<cr>
Example
PROG:DEL:ALL USB
Query Format
PROGram:CATalog? <cr>
Description
This query returns a comma separated list of <nr1> values
representing stored programs (e.g. 1,2,7…). If the MANUAL MODE
program setup exists, the list will include 0. If no programs are stored
at all, this query returns a 1
Parameters
Optional: Memory type: INTERNAL | RAM | USB
Parameter Format
<cr>
Returned Data Format
<nr1>,<nr1>,….,<nr1>,<nr1>
Query Example
PROG:CAT?
0,1,2,9,12
PROG:CAT? INTERNAL
0