28 | Ottobock
SensorHand Speed
Program 3
: 1 linear transducer AutoControl (green coding plug)
Open
Close
Initial Grip
Strength
AutoGrasp Grip
Force
FlexiGrip Function
High speed pull on
linear transducer
Speed:
constant 300 mm/s
Very slow pull release on
linear transducer
Hand stops in open
position
Quick pull release on
linear transducer
Hand closes
Speed:
constant 300 mm/s
10 N
up to max. 100 N
effective at maximal
grip force
9.3.3 Control with 1 switch:
Open:
at maximum speed, as long as switch is activated
Close:
after releasing the switch the Hand automatically closes at maximum speed and
begins to grasp the object with 10 N grip force.
Attention
:
Plug the red/white cable of the 13E99 Connecting Cable to closing side of the
9E169 Coaxial Plug (Fig. 6: (1) Battery Connecting Cable, (2) Electrode, (3)
red/white wire of 13E99 Connecting Cable )
Program 3:
1 Switch AutoControl (green coding plug)
OPEN
CLOSE
Initial Grip
Strength
AutoGrasp Grip
Force Increase
FlexiGrip Function
Opens as long as
switch is activated.
Speed: constant
300 mm/s
Closes automatically
as soon as switch is
released
Speed: constant
300 mm/s
10 N
up to max. 100 N
effective at maximal
grip force
9.4 Program 4:
VarioControl
Blue coding plug
Control with 1 electrode or 1 linear transducer
With this program, the opening speed is determined by the level and speed of the muscle sig-
nal. The closing speed is determined by the reduction of the muscle tension.
FlexiGrip is effective at maximal grip force. Once the grasp is secure the SensorHand Speed
reverts to the initial grip force.
Open:
Proportional. Opening speed is determined by level and speed of muscle
tension
.
Close:
Proportional. Closing speed is determined by speed and strength of muscle
relaxation.
This also affects the level of the maximal AutoGrasp grip force
increase.
Stop:
Very slow muscle relaxation to the electrode, Hand stops in an opened position.
Example 1:
After opening the amputee very slowly relaxes the muscle. Closing is
simultaneous to the length of muscle relaxation time. The object is grasped with
low force (10 N). There is no automatic grip force increase.
Summary of Contents for SensorHand 8E38 8
Page 2: ...2 Ottobock SensorHand Speed 1 2 2 3 4 5 6 1 7 8...
Page 201: ...SensorHand Speed Ottobock 201 13 Ottobock Derma Clean 453H10 1 14 CE 93 42 IX Ottobock VII 15...
Page 218: ...218 Ottobock SensorHand Speed 14 93 42 IX I Ottobock VII 15...
Page 223: ...SensorHand Speed Ottobock 223 9 1 DMC 2 1 1 1 3 1 4 1 1...
Page 232: ...232 Ottobock SensorHand Speed 12 GmbH 13 453H10 1 14 CE IX VII 15...
Page 233: ......
Page 234: ......
Page 235: ......