52
GN Otometrics A/S
Tests
AccuScreen PRO User Manual
6.2.3
DPOAE Diagnostic tests
The LED flashes green during the test.
If the LED is red, this indicates an error during testing:
•
The probe is unstable, or
•
Artifacts are present, indicating high environmental noise or a
noisy baby.
In these cases, check the test conditions and make sure that the probe
fits.
If the probe does not fit correctly, the screen displays the word
LEAK
.
Check the probe.
The displayed DPOAE Diagnostic graph
The response or lack of response is shown in the probability circle
during testing.
A response is always in phase. When a response is registered, it is
mapped from the center towards the outer edge of the probability
circle, forming a more or less straight line. A
PASS
is determined
with a confidence of 99.7% for all frequencies.
Noise (= no response) is randomly distributed in the circle and will
never reach the outer edge.
If mapping is randomly distributed, the test is heading towards a
REFER
. Check the probe fit and the test environment, and make sure
that the patient is calm.
If required, press
STOP
to stop the test.
Tests with a
Stopped
result are not saved. Accordingly, they cannot be
retrieved in the
View results
screens and they cannot be printed
out.
If the message
CalFail
appears, refit the probe and make sure that it
fits correctly. Restart the test.
Summary of Contents for Madsen AccuScreen PRO
Page 8: ...User Manual AccuScreen PRO 8 GN Otometrics A S ...
Page 12: ...12 GN Otometrics A S Introduction AccuScreen PRO User Manual ...
Page 20: ...20 GN Otometrics A S Test Methods AccuScreen PRO User Manual ...
Page 26: ...26 GN Otometrics A S The AccuScreen PRO AccuScreen PRO User Manual ...
Page 32: ...32 GN Otometrics A S Before You Start AccuScreen PRO User Manual ...
Page 58: ...58 GN Otometrics A S Tests AccuScreen PRO User Manual ...
Page 60: ...60 GN Otometrics A S Quick Test AccuScreen PRO User Manual ...
Page 70: ...70 GN Otometrics A S The Transfer Delete Data Menu AccuScreen PRO User Manual ...
Page 86: ...86 GN Otometrics A S Maintenance AccuScreen PRO User Manual ...
Page 98: ...98 GN Otometrics A S User Manual AccuScreen PRO ...