GN Otometrics A/S
53
AccuScreen PRO User Manual
Tests
6.2.4
ABR tests
If required, press
PAUSE
to suspend the test. The test can be sus-
pended for max. 2 minutes. Press
CONT
to resume the test.
Press
PAUSE
if the test progresses well but is suddenly influenced
by noise or movements.
Press
PAUSE
to store data registered up to this point in time. This
will prevent this data to be influenced by noise or movements.
If required, press
STOP
to stop the test. If you do, the test data will
not be stored.
6.3
After the test
6.3.1
Adding comments to a test
If you select
Add comments
, the
COMMENTS
screen appears.
You can either select from a list of 25 standard comments or add a
comment of your choice. You can edit the 25 standard comments
from AccuLink and then download them to the AccuScreen PRO.
Selecting from the list of comments
•
Scroll to select the appropriate comment and press
OK
to store
the selected comment.
Adding a new comment
•
Press
ADD
in the
COMMENTS
screen and use the alphanu-
meric keypad to enter a comment with a wording of your
choice of up to 21 characters. Press
OK
to store the comment.
Viewing comments
1.
You can view the stored comment and the test results relating
to the selected patient. To do so, select
View results
. See also
Chapter 9, “View Results” on page 65
.
Summary of Contents for Madsen AccuScreen PRO
Page 8: ...User Manual AccuScreen PRO 8 GN Otometrics A S ...
Page 12: ...12 GN Otometrics A S Introduction AccuScreen PRO User Manual ...
Page 20: ...20 GN Otometrics A S Test Methods AccuScreen PRO User Manual ...
Page 26: ...26 GN Otometrics A S The AccuScreen PRO AccuScreen PRO User Manual ...
Page 32: ...32 GN Otometrics A S Before You Start AccuScreen PRO User Manual ...
Page 58: ...58 GN Otometrics A S Tests AccuScreen PRO User Manual ...
Page 60: ...60 GN Otometrics A S Quick Test AccuScreen PRO User Manual ...
Page 70: ...70 GN Otometrics A S The Transfer Delete Data Menu AccuScreen PRO User Manual ...
Page 86: ...86 GN Otometrics A S Maintenance AccuScreen PRO User Manual ...
Page 98: ...98 GN Otometrics A S User Manual AccuScreen PRO ...