42
GN Otometrics A/S
Tests
AccuScreen PRO User Manual
If the result is a
PASS
, a
PASS
message appears before the detailed
results are shown.
The following result details are shown:
Time
Duration. Max. time depends on environmental
noise conditions
TEOAE
Significance of TEOAE signals.
A
Artifact rate
S
Stimulus stability rate
The graphic display shows the peaks reaching the ±3 sigma criteri-
on. Note that the graph is a statistical graph.
If the result is a
REFER
, the detailed results are shown immediately
after the test.
•
Add comments and view results as required. See
Section 6.4, ‘Interpretation of
•
You can now scroll to select the baby’s other ear on the screen
and, if required, reposition the baby so that you can test the
other ear.
•
Before you continue testing the other ear, check to make sure
that the probe is clean. See also
6.1.2
Test process DPOAE
Applies to AccuScreen PRO D, DA, TD, TDA.
1.
Do as described in
Chapter 5, “Preparing for testing” on page
PASS
Summary of Contents for Madsen AccuScreen PRO
Page 8: ...User Manual AccuScreen PRO 8 GN Otometrics A S ...
Page 12: ...12 GN Otometrics A S Introduction AccuScreen PRO User Manual ...
Page 20: ...20 GN Otometrics A S Test Methods AccuScreen PRO User Manual ...
Page 26: ...26 GN Otometrics A S The AccuScreen PRO AccuScreen PRO User Manual ...
Page 32: ...32 GN Otometrics A S Before You Start AccuScreen PRO User Manual ...
Page 58: ...58 GN Otometrics A S Tests AccuScreen PRO User Manual ...
Page 60: ...60 GN Otometrics A S Quick Test AccuScreen PRO User Manual ...
Page 70: ...70 GN Otometrics A S The Transfer Delete Data Menu AccuScreen PRO User Manual ...
Page 86: ...86 GN Otometrics A S Maintenance AccuScreen PRO User Manual ...
Page 98: ...98 GN Otometrics A S User Manual AccuScreen PRO ...