If you are doing only one measurement, select
Select Single
ear
. Press
OK
to confirm your selection.
9.
Select the ear you wish to test. Press
OK
to confirm your selec-
tion.
Impedance testing
Before the ABR test starts, the AccuScreen PRO carries out an im-
pedance test.
White
indicates the skin impedance of the white electrode (high on the
forehead)
Red
indicates the skin impedance of the red electrode (the nape of the
neck).
Less than 6000
Ω
:
Good
6000 to 12000
Ω
:
Fair
More than 12000
Ω
:
Poor (testing is not possible,
OK
is
not available, only
STOP
can be
pressed).
"Balance"
indicates the difference in the skin impedance of the white and the
red electrode. This value should also be fairly low (> 4000
Ω
):
0 to 1999
Ω
:
Good
2000 to 3999
Ω
:
Fair
More than 4000
Ω
:
Poor (testing is not possible,
OK
is
not available, only
STOP
can be
pressed).
Summary of Contents for Madsen AccuScreen PRO
Page 8: ...User Manual AccuScreen PRO 8 GN Otometrics A S ...
Page 12: ...12 GN Otometrics A S Introduction AccuScreen PRO User Manual ...
Page 20: ...20 GN Otometrics A S Test Methods AccuScreen PRO User Manual ...
Page 26: ...26 GN Otometrics A S The AccuScreen PRO AccuScreen PRO User Manual ...
Page 32: ...32 GN Otometrics A S Before You Start AccuScreen PRO User Manual ...
Page 58: ...58 GN Otometrics A S Tests AccuScreen PRO User Manual ...
Page 60: ...60 GN Otometrics A S Quick Test AccuScreen PRO User Manual ...
Page 70: ...70 GN Otometrics A S The Transfer Delete Data Menu AccuScreen PRO User Manual ...
Page 86: ...86 GN Otometrics A S Maintenance AccuScreen PRO User Manual ...
Page 98: ...98 GN Otometrics A S User Manual AccuScreen PRO ...