GN Otometrics A/S
43
AccuScreen PRO User Manual
Tests
2.
Select
DPOAE
from the
SELECT TEST
menu and select the ear
on which you wish to start the test.
3.
Press
OK
to start the test.
Calibration
Calibration is performed before the actual test starts.
Probe fit
If the probe does not fit correctly, the screen displays the word
LEAK
.
Adjust the position of the probe so that the graphical line between
1K and 2K is as horizontal as possible.
Balance test
The probe fit is followed by a balance test to make sure that the two
microphones function correctly.
The test starts automatically when calibration is completed.
The DPOAE test
The test as it progresses is shown on the display. See also
6.2.2, ‘DPOAE tests” on page 51
for monitoring the test.
Test duration
Approximately 10 seconds on patients with normal hearing.
The test frequencies are shown at the top of the display. The se-
quence of the test frequencies is as follows:
1.
4000Hz
2.
3200Hz
3.
2500Hz
4.
2000Hz
If required, press
SKIP
to skip testing of certain frequencies. The test
continues with the next frequency.
Summary of Contents for Madsen AccuScreen PRO
Page 8: ...User Manual AccuScreen PRO 8 GN Otometrics A S ...
Page 12: ...12 GN Otometrics A S Introduction AccuScreen PRO User Manual ...
Page 20: ...20 GN Otometrics A S Test Methods AccuScreen PRO User Manual ...
Page 26: ...26 GN Otometrics A S The AccuScreen PRO AccuScreen PRO User Manual ...
Page 32: ...32 GN Otometrics A S Before You Start AccuScreen PRO User Manual ...
Page 58: ...58 GN Otometrics A S Tests AccuScreen PRO User Manual ...
Page 60: ...60 GN Otometrics A S Quick Test AccuScreen PRO User Manual ...
Page 70: ...70 GN Otometrics A S The Transfer Delete Data Menu AccuScreen PRO User Manual ...
Page 86: ...86 GN Otometrics A S Maintenance AccuScreen PRO User Manual ...
Page 98: ...98 GN Otometrics A S User Manual AccuScreen PRO ...