For DMF configuration and calibration refer to your M2011 SiteSentinel
®
Nano
®
Configuration Guide.
You will need the A and B Factors that are etched into the body of the float. See the image below.
NOTICE:
The Density Measurement Float is not to be used in a pressurized tank.
8.2 Tank Thresholds
l
Make sure that the Density float is installed between the Water Float and the Product Float. When you
set the product threshold levels (
Low
and
Low-Low
) in the console's
Tank Thresholds
screen, make
sure the values are higher than where the Density Float is installed. Refer to your console's
configuration guide.
l
Measure the distance between the end of the probe shaft and the top end of the Density Float. Add 5.1
cm (2 inches) to this distance to include the dead zone at the end of the probe. The result is the value of
the minimum
Low-Low
product threshold.
Calculate Low-Low Threshold
Doc. No.: M2010-EU Rev.: 13
Page 59 of 149
Summary of Contents for M2010-EU
Page 6: ...Warranty 147 Doc No M2010 EU Rev 13 Page 6 of 149...
Page 20: ...Main Board Connections Jumpers and LEDs Doc No M2010 EU Rev 13 Page 20 of 149...
Page 21: ...Conduit Installation Doc No M2010 EU Rev 13 Page 21 of 149...
Page 22: ...Field Wiring Diagram Doc No M2010 EU Rev 13 Page 22 of 149...
Page 32: ...Doc No M2010 EU Rev 13 Page 32 of 149...
Page 38: ...924B Probe Installation DMP Probe Installation Doc No M2010 EU Rev 13 Page 38 of 149...
Page 41: ...Doc No M2010 EU Rev 13 Page 41 of 149...
Page 86: ...Typical Installation Drawings 30 0230 S Doc No M2010 EU Rev 13 Page 86 of 149...
Page 110: ...Appendix C Declaration of Conformity Doc No M2010 EU Rev 13 Page 110 of 149...
Page 111: ...Doc No M2010 EU Rev 13 Page 111 of 149...
Page 112: ...Doc No M2010 EU Rev 13 Page 112 of 149...
Page 113: ...Doc No M2010 EU Rev 13 Page 113 of 149...
Page 142: ...Doc No M2010 EU Rev 13 Page 142 of 149...
Page 146: ...Doc No M2010 EU Rev 13 Page 146 of 149...
Page 149: ......