3-19
ZUV
User’s Manual
Sect
ion 3
SETUP
Section 3
Setting Cumulative Alarms
Setting Cumulative Alarms
The cumulative energy of each head is stored on controllers. The head life can be judged
by setting thresholds to this energy value.
An alarm is displayed and an error is output when the cumulative energy of the head
exceeds the threshold. (default value: OFF)
■
Setting the cumulative alarm output threshold
LOCK mode-[TOOL]-[ALARM SET]-[1CH to 4CH]-[ON/OFF]
• Setting of cumulative energy is enabled only in the READY mode. Irradiated energy is not
accumulated in the LOCK mode since irradiation is not performed.
• When LOCK mode- [SYSTEM] - [OPTION] - [LOG DATA] - [OFF] is selected, the cumulative alarm
setting does not function.
Setting ready signals output conditions p.3-21
• After the head is replaced, initialize the cumulative energy value in the LOCK mode.
Initializing Setting Data p.3-22
Setting value
Description
ON
Enables setting of the cumulative energy threshold.
Range: 0.0 to 99999999.9
OFF
Disables setting of the cumulative energy threshold.
HEAD LONGEVITY
1CH
Cumulative alarm display