3-10
Section 3
Setting Value Obtaining and Changing Command
ZS Series
CompoWay/F Communication Command Reference
Section 3
Unit No.
and P
a
rameter No.
02h 02h 0000h
It indicates a normal end.
■
Measurement Setting 2 (common to all TASKs)
Unit No.
Data No.
Parameter
Setting range/
output range
Remarks
78h 00h
Non-measurement
setting
0: Keep
1: Clamp
78h
01h
Output upon clamp
0: MAX
1: 20mA
2: 12mA
3: 4mA
4: MIN
79h
00h
HYS
0 to 999999999 (nm)
79h
01h
Timer mode
0: OFF
1: OFF DELAY
2: ON DELAY
3: 1 shot
79h
02h
Delay time
1 to 5000 (ms)
7Ah
02h
Monitor focus mode
0: OFF, 1: ON
(Note 4)
7Ah 03h
Monitor
focus
distance value 1
-999999999 to 999999999 (nm)
7Ah 04h
Monitor
focus
distance value 2
-999999999 to 999999999 (nm)
7Ah
05h
Monitor focus current
value 1
4 to 20 (mA)
7Ah
06h
Monitor focus current
value 2
4 to 20 (mA)
7Ah
07h
Monitor focus voltage
value 1
-10 to 10 (V)
7Ah
08h
Monitor focus voltage
value 2
-10 to 10 (V)
7Eh
04h
Input polarity of
external input 0
7Eh
05h
Input polarity of
external input 1
7Eh
06h
Input polarity of
external input 2
7Eh
07h
Input polarity of
external input 3
0: L active
1: H active
7Fh 01h
External
Input
(IN) function mode
0: Standard
1: Bank
7Fh
06h
Digital output target
0: OFF
1: ON
7Fh
0Ah
External input 0 mode
7Fh
0Bh
External input 1 mode
(Note 5)
7Fh
0Ch
External input 2 mode
7Fh
0Dh
External Input 3 mode
0: Not used
1: Trigger
2: Hold reset
3: Laser OFF
4: Zero-reset
F0h
08h
External input mode
0: STAND
2: Parallel input OFF mode
(Note 6)