Setting Value Obtaining and Changing Command
ZS Series
CompoWay/F Communication Command Reference
Section 3
Unit No.
and P
a
rameter No.
Unit No.
Data No.
Parameter
Setting range/
output range
Remarks
F0h
C0h
Timing input
0: OFF
1: ON
F0h
C1h
Reset input
0: OFF
1: ON
F0h
C2h
LD-OFF input
0: OFF
1: ON
F0h
C3h
Zero-reset execution
1: Execution
F0h
C4h
Zero-reset cancel
1: Cancel
(Note 4)
To change the monitor focus setting, monitor focus mode must be ON.
For example, to change the current setting of the monitor focus, change parameters below:
-
Monitor focus mode (ON)
-
Monitor focus distance value 1
-
Monitor focus distance value 2
-
Monitor focus current value 1
-
Monitor focus current value 2
In addition, set outputs (current and voltage), which are set by the analog output switch in the controller.
(Note 5)
External input 0 mode and External input 1 mode settings cannot be changed
because they automatically function as bank inputs when the bank mode is used (when “external input
(IN) function mode” is set to “bank”).
(Note 6)
When the external input mode is set to “STAND”, the controller enters into the status below:
*
External input of the controller is enabled,
*
Commands from timing input to zero-resetting are disabled.
When the external input mode is set to “Parallel input OFF”, the controller enters into the status below:
*
External input of the controller is enabled,
*
Commands from timing input to zero-resetting are disabled.
Therefore, to send a command related to inputs, the external input mode should be set to “Parallel input
OFF.”
Once ON is written to the timing input, reset input, and LD-OFF input parameters, the ON state is kept
until OFF is written next.
Controller operates in the same manner as the input state is kept ON.
However, special care shall be taken because the operation changes depending on the state of external
input line to the controller when the external input mode is set back to the “standard mode.”
Operations are performed depending on the settings when “Execution” or “Cancel”
is written to the parameter with respect to the zero-reset.
Example 1: To perform zero-reset (when ZS-LDC is 1CH):
The command sending sequence is as follows:
(1)
Write “parallel input OFF mode” to the “external input mode” parameter.
(2)
Write “execution” to the “zero-reset execution” parameter.
(3)
Write “STAND” to the “external input mode” parameter.