3-17
Section 3
Setting Value Obtaining and Changing Command
ZS Series
CompoWay/F Communication Command Reference
Section 3
Unit No.
and P
a
rameter No.
Unit No.
(Note 1)
Data No.
Parameter
Setting range/
output range
Remarks
2Bh 02h
Average number
of times
0: 1
1: 2
2: 4
3: 8
4: 16
5: 32
6: 64
7: 128
8: 256
9: 512
10: 1024
11: 2048
12: 4096
2Ch 02h Differentiation
mode
0: OFF
1: ON
2Ch
03h
Differentiation cycle
1 to 5000 (ms)
2Dh 02h Hold
mode
0: OFF
1: PEAK
2: BOTTOM
3: P-P
4: AVERAGE
5: SAMPLE
2Dh
03h
Trigger method
0: External
1: Self-up trigger
2: Self-down trigger
2Dh
04h
Trigger level
-999999999 to 999999999
(nm)
2Dh
05h
Trigger hysteresis
0 to 999999999 (nm)
2Dh
06h
Trigger delay (ms)
0 to 5000 (ms)
2Dh
07h
Sampling period (ms)
1 to 5000 (ms)
2Dh
08h
Trigger delay mode
0: OFF
1: ON
2Eh
05h
Offset value for zero r
eset
-999999999 to 999999999
(nm)
2Eh 07h
Zero reset mode
0: REAL
1: HOLD
30h 02h
LOW threshold value
-999999999 to 999999999
(nm)
30h 03h
HIGH threshold value
-999999999 to 999999999
(nm)
(Note 1)
The Unit No. here indicates TASK 1 Unit No.
Unit No. other than the above, such as TASK2 or a later one, can be calculated
by adding (TASK number -1) to the Unit No. and multiplying it by 14h for another reference or setting.