How to Obtain Flow Data
ZS Series
CompoWay/F Communication Command Reference
Section 4
Ho
w to
Obt
a
in Flo
w
Da
ta
Setting Parameters
This section describes buffer size, buffer interval, and logging data.
Buffer
Size
and
Buffer
Interval
Buffer size of the measurement data to be accumulated in the ZS controller and its buffer
interval can be changed. To change buffer size and buffer interval, change the parameters
below:
* Use DATAGET and DATASET commands to change and view the parameter.
Unit No.
Data No.
Parameter
Details
7Ch 2
Flow data accumulation
mode
0: OFF
1: ON
7Ch 3
Buffer interval
1 to 65535
7Ch 4
Buffer size
1 to 1000
Logging
Data
Type
Data to be accumulated in the ZS controller can be selected.
■
When using ZS-LDC
Up to 3 data types can be accumulated.
Unit No.
Data No.
Parameter
Details
7Ch 5
Accumulation data 1
7Ch 6
Accumulation data 2
7Ch 7
Accumulation data 3
*
* Setting varies depending on the measurement mode.
- When “THICK/GAP - FILM/OTHERS - TRANSFER” mode is selected
Value
Details
0 No
accumulation
1
Result of area 1
2
Result of area 2
3 Thickness/gap
value
Setting Parameters