Setting Value Obtaining and Changing Command
ZS Series
CompoWay/F Communication Command Reference
Section 3
Unit No.
and P
a
rameter No.
Parameter List (ZS-MDC)
■
Input Setting (common to all TASKs)
Unit No.
Data No.
Parameter
Setting range/output range
Remarks
00h 00h
Data A input
destination CH
00h 01h
Data B input
destination CH
00h 02h
Data C input
destination CH
00h 03h
Data D input
destination CH
00h 04h
Data E input
destination CH
00h 05h
Data F input
destination CH
00h 06h
Data G input
destination CH
00h 07h
Data H input
destination CH
00h 08h
Data I input
destination CH
0 to 11 (CH)
00h
09h
Data A input mode
00h
0Ah
Data B input mode
00h
0Bh
Data C input mode
00h
0Ch
Data D input mode
00h
0Dh
Data E input mode
00h
0Eh
Data F input mode
00h
0Fh
Data G input mode
00h
10h
Data H input mode
00h
11h
Data I input mode
0: OFF
1: ON
00h
20h
Obtained result A
00h
21h
Obtained result B
00h
22h
Obtained result C
00h
23h
Obtained result D
00h
24h
Obtained result E
00h
25h
Obtained result F
00h
26h
Obtained result G
00h
27h
Obtained result H
00h
28h
Obtained result I
-999999999 to 999999999
(nm)
Read only
Example 1: When assigning (writing) input A to CH3:
To assign input A to CH3, input destination CH of data A should be set to 3CH and input mode of data A
should be ON.
(1) For input destination CH of data A, both Unit No. and Data No. are 00h.
(2) For input mode of data A, Unit No. and Data No. are 00h and 09h, respectively.
Therefore, commands to be sent are shown as below:
(1) When 02h 02h (C000h + 00h (Data No.)) 00h(Unit No.) 00h (CH No.),