Getting Started
13
2
Enter target blood glucose value
1. Use the
Up/Down Controller
buttons to enter a
target blood
glucose
(BG) value, then press
Next
.
2. Use the
Up/Down Controller
buttons to enter the “correct
above” value or
correction threshold
(the BG value above
which you want the PDM to suggest a
correction bolus
), then
press
Next
.
3. To save the target BG value, press
Done
. To change these set-
tings, select the segment you wish to change and press
Edit
.
You can define up to 4 target BG time segments. To enter
additional segments, choose
[add new]
, then:
a. Press
New
.
b. Enter a start time for the target BG segment, then press
Enter
.
c. Enter an end time for the target BG segment, then press
Next
.
d. Enter a target blood glucose value, then press
Next
.
e. Use the
Up/Down Controller
buttons to enter the “cor-
rect above” value or correction threshold (the BG value
above which you want the PDM to suggest a correction
bolus), then press
Next
.
f. When you have finished entering segments, press
Done
.
4. To see the segments as a list, press
List
. To return to the
graphic view, press
Graph
.
5. Press
Save
to accept the target BG profile you have set.
Enter minimum BG allowed for use in bolus calculation
1. Use the
Up/Down Controller
buttons to enter a minimum
blood glucose value, below which the System will not calcu-
late a suggested bolus. The values are in 1 mg/dL increments
and range from 50 to 70 mg/dL.
2. Press
Next
.
Enter insulin-to-carbohydrate ratio
1. Use the
Up/Down Controller
buttons to enter your
insulin-
to-carbohydrate ratio
, then press
Next
. (Insulin-to-carbohy-
drate ratio is the number of carbs covered by 1 unit of insulin.)
If the suggested bolus calculator is
On
, a correction
bolus will be suggested only when your BG is above
the value shown in brackets on the screen.