Errors, Advisories, and Hazard Alarms
128
10
Pod deactivation required.
Press “Ok” to deactivate.
PDM
Press
OK
to acknow-
ledge.
Change Pod.
Check blood glucose.
Pod Error.
Insulin delivery stopped.
Change pod now.
Pod
or
PDM
Press
OK
to acknow-
ledge.
Change Pod.
Check blood glucose.
PDM Error.
Remove pod now.
Call customer support:
800.591.3455
Ref: 05-100-0618-00000
PDM
Press
OK
to acknow-
ledge.
Remove Pod.
Call Customer Care at
800-591-3455 (from
outside the United
States, 781-457-5098).
Check blood glucose.
Alarm
Source Action
PDM Error.
Press “Ok” to reset clock.
PDM
Press
OK
to acknow-
ledge.
Reset clock.
Check blood glucose.
PDM Error.
Memory corruption.
Press “Ok” to reset PDM
and delete all user set-
tings. History records will
not be deleted.
PDM
Press
OK
to acknow-
ledge.
Change Pod.
Check blood glucose.
System Error.
Remove pod now.
Call customer support:
800.591.3455
Ref: 05-100-0618-00000
PDM
or
Pod
Press
OK
to acknow-
ledge.
Remove Pod.
Call Customer Care at
800-591-3455 (from
outside the United
States, 781-457-5098).
Check blood glucose.
Alarm
Source Action