Page: 145
Type Code
Input Range
Data Format
+F.S
-F.S
0A -1
~
+1
V
Engineering Unit
+10000 -10000
2’s comp HEX
7FFF 8000
0B
-500 ~ +500 mV
Engineering Unit
+5000
-5000
2’s comp HEX
7FFF
8000
0C
-150 ~ +150 mV
Engineering Unit
+15000
-15000
2’s comp HEX
7FFF
8000
0D
-20 ~ +20 mA
Engineering Unit
+20000
-20000
2’s comp HEX
7FFF
8000
1A
0 ~ +20 mA
Engineering Unit
+20000
0
2’s comp HEX
FFFF
0000
0E
Type J
Thermocouple
-210 ~ 760°C
Engineering Unit
+7600
-2100
2’s comp HEX
7FFF
DCA2
0F
Type K
Thermocouple
-270 ~ 1372°C
Engineering Unit
+13720
-2700
2’s comp HEX
7FFF
E6D0
10
Type T
Thermocouple
-270 ~ 400°C
Engineering Unit
+4000
-2700
2’s comp HEX
7FFF
A99A
11
Type E
Thermocouple
-270 ~ 1000°C
Engineering Unit
10000
-2700
2’s comp HEX
7FFF
DD71
12
Type R
Thermocouple
0 ~ 1768°C
Engineering Unit
+17680
0
2’s comp HEX
7FFF
0000
13
Type S
Thermocouple
0 ~ 1768°C
Engineering Unit
+17680
0
2’s comp HEX
7FFF
0000