
Page: 116
The calibration procedure is as follows:
Step 1 Warm up the module for latest 30 minutes.
Step 2 Set the type code to the type you wish to calibrate.
Step 3 Enable calibration.
Step 4 Connect the zero calibration resistor.
(Refer to the table
–
Type Code)
Step 5 Send the zero calibration command.
(Refer to the table - Modbus Address for Calibration)
Step 6 Connect the span calibration resistor.
(Refer to the table
–
Type Code)
Step 7 Send the span calibration command.
(Refer to the table - Modbus Address for Calibration)
Step 8 Disable calibration.
Summary of Contents for OME-ET-7000 Series
Page 1: ......
Page 2: ......
Page 37: ...Page 32 The lower section of the page provides a brief introduction to the web interface...
Page 86: ...Page 81 After completing the configuration click the Submit button to save the settings 1 2...
Page 92: ...Page 87 The default URL that the More Information item will link to http www omega com...
Page 158: ......
Page 159: ......