V
DC volts
A
DC current
OVER
measurement is over or under
range
or
one of the pressure units listed
with the ISO_PRES_UNIT?
command
and where <primary-value> is the present
primary measurement expressed in scientific
notation.
and where <primary-units> is one of the
following:
FAR
°F
CEL
°C
OHM
Ohms
V
DC volts (thermocouple millivolts)
OVER
measurement is over or under
range
NONE primary display is presently set to
a source mode
or
one of the pressure units listed
with the PRES_UNIT? command
Example:
VAL?
2.00,V,5.01,CEL
This example indicates that the isolated
measurement is 2.137 volts and that the primary
measurement is 57.633 °C.
*WAI
IEEE-488
RS-232
Sequential
Overlapped
This command prevents further remote commands from being
executed until all previous remote commands have been completed.
For example, if you send an OUT command, you can cause the
CL3001 to wait until the output has settled before continuing on to
the next command if you follow OUT with a *WAI command. The
*WAI command is useful with any overlapped command, preventing
the CL3001 from processing other commands until the overlapped
command is completed.
Parameter:
<None>
Response:
<None>
Example:
OUT 1.1 V
*WAI
X
X
X
84