![background image](http://html1.mh-extra.com/html/omega/cl3001/cl3001_user-manual_737625010.webp)
Item
Name
Description
1
VOLTS
DC voltage output terminals.
See notes 1 and 2 below.
2
mA
DC current output terminals.
See notes 1 and 2 below.
3
RTD/O OUTPUT
Two wire RTD and Ohms output terminals.
See notes 1 and 2 below.
4
TC INPUT/OUTPUT
Thermocouple input and output terminals.
These terminals accept a miniature
polarized thermocouple plug with flat in-line
blades spaced 7.9 mm (0.312 in.) center to
center.
5
4W RTD/O INPUT
Four wire RTD and Ohms input terminals.
See notes 1 and 3 below.
6
Pressure module input connector.
Note 1:
These terminal binding posts are made of a special copper alloy
to reduce thermal EMF's. They support the use of either discreet
wires or standard banana plugs, and the HI/LO pairs are spaced
for standard dual banana plugs.
Note 2:
Caution
. Do not exceed a maximum of 100 volts to chassis
ground.
Note 3:
Caution
. Do not exceed a maximum of 20 volts to chassis
ground.
2.3 Primary Input/Output Display and Controls
Figure 3 describes the primary input/output display and controls in
detail.
Figure 3 - Primary Input/Output Display and Controls
6