FUNC?
IEEE-488
RS-232
Sequential
Overlapped
This command returns the present output, measurement, or
calibration function for the primary and isolated displays.
Parameter:
<None>
Response:
<isolated>,<primary>
where <isolated> is one of the following:
DC10V
measure DC voltage, 10V
range
DC100V
measure DC voltage,
100V range
DCI
measure DC current
PRESSURE measure
pressure
and where <primary> is one of the following:
DCV
source DC voltage
DCI
source DC current
RTD_OUT
source RTD/Ohms
RTD_IN
measure RTD/Ohms
TC_OUT
source thermocouple
TC_IN
measure thermocouple
PRESSURE measure
pressure
Example:
FUNC?
DC10V,PRESSURE
This example indicates that the isolated display is
selected to the 10V range and the primary display
is selected to pressure.
HART?
IEEE-488
RS-232
Sequential
Overlapped
This command returns the isolated DC current input HART resistor
status.
Parameter:
<None>
Response:
<value>
where <value> is ON or OFF
Example:
HART?
OFF
This example indicates that the isolated DC
current input HART resistor is turned off.
X
X
X
X
X
X
64