LOOP_POWER_OFF
IEEE-488
RS-232
Sequential
Overlapped
This command disables the isolated DC current input 24V loop
power.
Parameter:
<None>
Response:
<None>
Example:
LOOP_POWER_OFF
This example disables the isolated DC current
input 24V loop power.
LOOP_POWER_ON
IEEE-488
RS-232
Sequential
Overlapped
This command enables the isolated DC current input 24V loop
power.
Parameter:
<None>
Response:
<None>
Example:
LOOP_POWER_ON
This example enables the isolated DC current
input 24V loop power.
*OPC
IEEE-488
RS-232
Sequential
Overlapped
Operations Complete command. This command sets bit 0 (OPC) of
the Event Status Register to 1 when all pending device operations
are complete. Also see the *ESR? command.
Parameter:
<None>
Response:
<None>
Example:
*OPC
This example sets bit 0 of the Event Status
Register to 1 when all pending device operations
are done.
*OPC?
IEEE-488
RS-232
Sequential
Overlapped
Operations Complete query. This command returns a 1 after all
pending CL3001 operations are complete. This command does not
respond until all pending CL3001 operations are complete, causing
the control program execution to pause until operations are
complete. Also see the *WAI command.
Parameter:
<None>
X
X
X
X
X
X
X
X
X
X
X
X
69