Status Commands
Command
Description
FAULT?
Returns the most recent error code in the CL3001 error
queue, and then removes that error code from the
queue.
10.3
Error Code Listing
Error Number
Message Class
Description
1
DDE
Error queue overflow.
100
EXE
OPER or STBY was received when the
CL3001 is in measure mode or thermocouple
source mode.
101
CME
A non-numeric entry was received in a field
that should contain a numeric entry.
102
EXE
The numeric field exceeds 10 characters.
103
CME
Invalid units name or prefix.
104
EXE
An attempt to enter RTD source mode was
made when SPRT is selected, or to select
SPRT when RTD source mode is selected.
105
EXE
Entry is above upper limit for the selected out-
put range.
106
EXE
Entry is below lower limit for the selected out-
put range.
108
CME
A required command parameter was missing.
109
CME
An invalid TC_MEAS or RTD_MEAS unit
parameter (not CEL or FAR) was received, or
an invalid PRES_UNIT or ISO_PRES_UNIT
parameter was received.
110
CME
An invalid RANGELCK parameter was
received.
111
EXE
RANGELCK ON was received when the
CL3001 is not in Volts mode.
112
CME
An invalid RTD_TYPE parameter was
received.
113
CME
An invalid TC_REF parameter was received.
114
CME
An invalid TSENS_TYPE parameter was
received.