![background image](http://html1.mh-extra.com/html/omega/cl3001/cl3001_user-manual_737625100.webp)
RTD and Thermistor Specification, Input (continued)
Absolute Uncertainty,
tcal ±5 °C, ±(°C)
1
Range (° C)
Output/Input
RTD Type
Minimum
Maximum
90 days
1 year
Pt 385, 1000
Ω
-200 °C
-80 °C
0.011 °C
0.012 °C
-80 °C
0 °C
0.014 °C
0.015 °C
0 °C
100 °C
0.019 °C
0.020 °C
100 °C
260 °C
0.025 °C
0.028 °C
260 °C
300 °C
0.027 °C
0.030 °C
300 °C
400 °C
0.030 °C
0.034 °C
400 °C
600 °C
0.041 °C
0.045 °C
600 °C
630 °C
0.042 °C
0.047 °C
Ni 120, 120
Ω
-80 °C
0 °C
0.009 °C
0.010 °C
0 °C
100 °C
0.010 °C
0.011 °C
100 °C
260 °C
0.011 °C
0.012 °C
Cu 427, 10
Ω
-100 °C
260 °C
0.067 °C
0.069 °C
YSI 400
15 °C
50 °C
0.005 °C
0.007 °C
SPRT
-200 °C
660 °C
0.05 °C
0.06 °C
1. 4-wire mode. Uncertainties listed do not include probe uncertainties