Example:
*OPT?
0
The example indicates that no options are
installed.
OUT
IEEE-488
RS-232
Sequential
Overlapped
This command sets the output mode and value of the CL3001. To
source a temperature, select the desired mode and sensor
parameters first with the TSENS_TYPE, RTD_TYPE, and TC_TYPE
commands. Use the multiplier prefixes k for kilo, m for milli, and u
for micro with the OUT command units, as desired. The unit may be
omitted if the output mode is not to be changed.
Parameter:
<value> <units>
where <value> is the output value to be
generated
and where <units> is one of the following, with a
possible prefix multiplier (k, m, or u) as described
above:
V
DC volts
A
DC current
OHM
Resistance
CEL
Temperature in Celsius
FAR
Temperature in fahrenheit
Response:
<None>
Examples:
Explanation
OUT 15.2 V
Change to DC volts, output 15.2 V
OUT 1.2 mA
Change to DC current, output 1.2
mA, note units prefix multiplier m
OUT 5 Ohm
Change to Resistance, output 5 O
OUT 100 CEL
Change to temperature in °C,
output 100 °C
OUT 3
No change to output mode,
output a value of 3 in the present
units
OUT?
IEEE-488
RS-232
Sequential
Overlapped
This command returns the present output value and units of the
CL3001.
Parameter:
<None>
Response:
<value>,<units>
X
X
X
X
X
X
71