Warning
. Automatic setpoints over 30V will not place the CL3001
in standby mode for each new value.
f)
The CL3001 can be locked to a specific voltage range by
entering a value in that range and then selecting the secondary
RNG LOCK function by pressing the
and
keys.
4.2 DC Current Output
The CL3001 can source DC current from 0 mA to 100 mA.
a) Disconnect any test leads from external devices.
b) Press the
key to select DC voltage and current mode, if
not already selected. If DC voltage mode is displayed, press the
key again to cycle to DC current mode.
c) Connect the unit under test to the current output terminals of the
CL3001 as shown in figure 11.
Figure 11 - DC Current Output Connection
d) Use the numeric keypad to enter the desired output value and
press the
key.
Alternatively, use the
or
cursor key to select a digit to
modify, followed by the
or
cursor key to ramp the digit
up or down. This method offers a simple solution when small
changes to an output value are required, or if specific decades
need to be incremented or decremented.
e) When DC current mode is first selected, the CL3001 is placed in
the standby (Stby) mode which puts the positive (+) output jack
into a high impedance state (>100k ohm) for safety. To place
the output into the active state, press the
key which
toggles between the standby and operate modes.