![background image](http://html1.mh-extra.com/html/omega/cl3001/cl3001_user-manual_737625096.webp)
Thermocouple Specification, Output and Input (continued)
Absolute Uncertainty,
tcal ±5 °C, ±(°C)
1
Range (° C)
Output/Input
TC Type
Minimum
Maximum
90 days
1 year
R
0 °C
250 °C
0.58 °C
0.58 °C
250 °C
400 °C
0.34 °C
0.35 °C
400 °C
1000 °C
0.31 °C
0.33 °C
1000 °C
1750 °C
0.30 °C
0.40 °C
S
0 °C
250 °C
0.56 °C
0.56 °C
250 °C
1000 °C
0.36 °C
0.36 °C
1000 °C
1400 °C
0.30 °C
0.37 °C
1400 °C
1750 °C
0.35 °C
0.46 °C
T
-250 °C
-150 °C
0.51 °C
0.63 °C
-150 °C
0 °C
0.18 °C
0.24 °C
0 °C
120 °C
0.13 °C
0.16 °C
120 °C
400 °C
0.12 °C
0.14 °C
U
-200 °C
0 °C
0.56 °C
0.56 °C
0 °C
600 °C
0.27 °C
0.27 °C
XK
-200 °C
-100 °C
0.22 °C
0.22 °C
-100 °C
300 °C
0.12 °C
0.13 °C
300 °C
800 °C
0.19 °C
0.20 °C
BP
0 °C
200 °C
0.42 °C
0.42 °C
200 °C
600 °C
0.32 °C
0.32 °C
600 °C
800 °C
0.39 °C
0.40 °C
800 °C
1600 °C
0.45 °C
0.46 °C
1600 °C
2000 °C
0.57 °C
0.58 °C
2000 °C
2500 °C
0.67 °C
0.80 °C
1. Does not include thermocouple wire error.
92
Summary of Contents for CL3001
Page 103: ......