116
EXE
No pressure module was present when a
pressure function was requested, or the total
zero offset from calibration is more than 6%.
This error can also occur if the total zero off-
set from calibration is out of limits when zero-
ing the thermocouple millivolts type (maxi-
mum offset ±1 mV), or the RTD ohms type
(maximum ±0.1 ohm on high range, or ±0.01
ohm on the low range).
117
CME
An unrecognizable command was received.
118
CME
An invalid parameter was received.
120
EXE
The serial input buffer overflowed
121
EXE
The command string buffer overflowed.
122
QYE
The serial output buffer overflowed.
123
DDE
The output overloaded. See display error
message OVER LOAD in section 2.7.
124
DDE
The CL3001 is out of tolerance. This error is
set after a failed initialization or a failed *TST?
command.
125
DDE
The CL3001 ADC has failed. This error is set
after a failed initialization or a failed *TST?
command.
61