3.
Remove the compartment cover. The fuses come out with
the compartment cover and can easily be checked or
replaced.
4.
To reinstall the fuse holder, push the compartment cover
back into the compartment until the tab locks in place.
11.3 Changing the Line Voltage
The calibrator arrives from the factory configured for the line voltage
appropriate for the country of purchase, or as specified when it is
ordered. To verify the line voltage setting, check the line voltage
indicator on the power line fuse compartment cover.
Confirm that the line voltage selection is set for 120 V for line
voltages between 90 V and 132 V, or that the selector is set to 240 V
for line voltages between 198 V and 264 V.
Warning
To avoid electrical shock hazard disconnect line power before
opening the case or fuse compartment.
To change the line voltage:
1.
Disconnect line power.
2.
Using the blade of a suitable flat screwdriver, pry up the
tab at the base of the line fuse compartment by inserting
the blade in the center slot under the tab. The
compartment cover will pop part way out.
3.
Remove the compartment cover.
4.
Remove the line voltage selector assembly by gripping the
line voltage indicator tab with pliers and pulling it straight
out of the compartment.
5.
Rotate the line voltage selector assembly to the desired
voltage and reinsert.
6.
Verify you are using the appropriate fuse for the selected
line voltage (see table 10 above), and reinstall the fuse
compartment by pushing it in until the tab locks in place.
87
Summary of Contents for CL3001
Page 103: ......