![OLS CASY TT Operator'S Manual Download Page 60](http://html1.mh-extra.com/html/ols/casy-tt/casy-tt_operators-manual_3812617060.webp)
60
CASY TT Operator‘s Guide
Format of Serial Data Transfer
Row Name of Parameter
Format
Comments
1 Serial Number
@@@@@@@@@@@@
2 Setup Name
@@@@@@@@@@
3 Setup Number
##
00 – 19
4 Time
##:##
24 h format
5 Date
dd.mm.yyyy or mm/dd/yyyy depending on the selected format
6 Dilution Factor
#.###e+##
7 Measuring Cycles
##
1 – 10
8 Sample Volume (µl)
###.##
corrected based on volume calibra-
tion
9 Mode of Aggregation Correction
#
0=Off, 1=Manual, 2=Auto
10 Manual Aggregation Correction
(fl)
#.###e+##
single cell volume
11 Percent Mode
#
0=Off, 1=Viab, 2=%Cnt, 3= %Vol
12 Debris Mode
#
0=Off, 1=On
13 Capillary Diameter (µm)
###
14 X-Axis Scale (µm)
###
15 Left Normalization Cursor (µm)
###.##
16 Right Normalization Cursor (µm)
###.##
17 Left Evaluation Cursor (µm)
###.##
18 Right Evaluation Cursor (µm)
###.##
19 Y-Axis
####
Auto or 10 – 7500
20
– 419
Counts in the 400 Size Channels
#####
420 Counts
##########
within evaluation cursor range
421 Concentration Range
#
0=Ok, 1=Too High
422 Counts > Upper Scaling Limit/ml
#.###e+##
423 Aggregation Factor
##.###
424 Counts/ml
#.###e+##
within evaluation cursor range*
425 Total Counts/ml
#.###e+##
within normalization cursor range*
426 Viable Cells/ml
#.###e+##
within evaluation cursor range*
427 Total Cells/ml
#.###e+##
within normalization cursor range*
428 Cell Debris/ml
#.###e+##
on left side of left normalization
cursor*
429 Volume/ml in fl
#.###e+##
within evaluation cursor range
430 Total Volume/ml in fl
#.###e+##
within normalization cursor range *
431 %Counts
###.##
based on normalization
range=100%*
432 %Viability
###.##
based on normalization
range=100%*
433 %Volume
###.##
based on normalization
range=100%*
434 Mean Volumen in fl
#.###e+##
within evaluation cursor range
435 Peak Volumen in fl
#.###e+##
within evaluation cursor range
436 Mean Diameter in µm
###.##
within evaluation cursor range
437 Peak Diameter in µm
###.##
within evaluation cursor range
438 Check Sum
XXXXXXXX
4 bytes hexadecimal
Depending on your setup settings, some data tagged with an asterisk (*) are output as
not
calculated
.
Data Output via the Serial Interface