CASY TT Operator‘s Guide
MAIN MENU – Edit Setup
SAVING SETUP
TT-2AB-9999 15.04.2009 10:15
Capillary: 150 µm
Setup No: 00 01 02 03 04 05 06 07 08 09
*
10 11 12 13 14 15 16 17 18 19
USE NEXT TO SELECT SETUP AND PRESS ENTER TO SAVE.
USE CANCEL TO WORK WITH NEW PARAMETERS WITHOUT SAVING.
You must save your entries if you want the settings to be available next time you switch on the CASY
System. To do this, press
NEXT
to select one of the 20 setups and then press
ENTER
. The setup cur-
rently loaded will be offered as a default setting. You may select both setups that have already been
used and tagged with * and also unused setups. If you select a setup tagged with *, its stored set-
tings will be overwritten with your new settings. If you select an unused setup, it will appear tagged
with * and will be available for selection next time
Select Setup
is opened.
If you do not wish to save your settings, press
CANCEL
. The new settings are used until you select
an alternative setup or switch off your CASY System. They will then no longer be available. As a setup
name for unsaved settings,
NOT SAVED
appears and the setup number is replaced by two forward
slashes //.
If you try to overwrite a logged setup, a warning message will appear. You will be reminded that
pressing
ENTER
will permanently delete this setup’s existing settings. Pressing
CANCEL
enables
you to return to the
Saving Setup
screen to select a free setup for saving.