A
A-101 Color Measurement Instruments
ES3640e MFP Color Guide - 101
5. Use the AC adapter to provide power.
Plug the small connector on the adapter cable into the power
input of the X-Rite DTP41 and plug the AC adapter into the power
cord. Plug the power cable into a wall outlet.
6. Turn on the computer.
7. Calibrate the X-Rite DTP41 using ColorWise Pro Tools (see the
following section).
8. Calibrate the Fiery using ColorWise Pro Tools and the DTP41 (see
Calibrating the DTP41
For the best color accuracy, calibrate the X-Rite DTP41
spectrophotometer every time you calibrate the Fiery. You can
initiate the calibration sequence in ColorWise Pro Tools Calibrator
as part of the process of Fiery calibration. The computer may also
prompt you that it is necessary to calibrate the
spectrophotometer.
Alternatively, you can initiate the DTP41 calibration process using
the Instrument button on the DTP41. For more information on this
method, see the documentation that came with the X-Rite DTP41.
AC adapter
Small connector
Power cord